The number of detectors, angular range, and layout each contribute to overall system performance.
The LA-960V2 is designed and built to provide the same experience regardless of manufacture date, operator skill, or geographic location. Achieve unmatched instrument agreement without the hassle of correlation.
Always make perfect measurements with computer-controlled laser alignment. The alignment process is finished in only a few seconds with HORIBA’s innovative approach.
The LA-960V2 is a highly refined particle size analyzer capable of accurately measuring NIST-traceable size standards within 0.6% of specification. Fully compliant with ISO 13320 recommendations regarding the measurement of materials on the D10, D50, and D90.
The LA-960V2 PowderJet combines several unique and patented features to provide the most reproducible dry measurements. Use the Auto Measurement function to control vacuum, air pressure, powder flow, start/stop conditions, measurement duration, and data processing. Designed to handle every application including small sample amounts, friable powders, and highly agglomerated materials.
| Measurement Principle | Mie scattering and Fraunhofer diffraction |
|---|---|
| Measurement Range | 10 nm – 5000 μm (Option dependent) |
| Measurement Time | Typical measurement takes 60 seconds from liquid filling, sampling and measurement to rinsing. |
| Measurement Method | Circulation measurement or fraction cell measurement (Fraction cell is optional) |
| Sample Quantity | Approximately 10 mg – 5 g (Depending on the particle size, distribution and density) |
| Dispersing Volume | Approximately 180 mL for standard pumping system, 5/10/15 mL for Fraction Cell accessory, Manual filling: 35 mL, Automatic filling: 40 mL for MiniFlow accessory. Approximately 1 L for LiterFlow option. |
| Compatible Liquids | A type: Water*, S type: Common organic solvents (*also mixtures with small volumes of ethanol) |
| Communication | USB 2.0 (USB 3.0 required for some options) |
| Light Sources | Red solid-state 5 mW laser diode (650 nm), Blue solid-state 3 mW LED (405 nm) |
| Dispersion System |
In-line ultrasonic probe: 30 W, 20 kHz, adjustable levels Circulation pump: Fully automated fill and circulation pumps, 15 adjustable speeds, 4 selectable fill levels, 15 selectable circulation speeds (max: 10 L/min) |
| Operating Conditions | 15–35°C (59–95°F), relative humidity 85% or less (no condensation) |
| Power | AC 100–240 V, 50/60 Hz, 300 VA |
| Dimensions | 705 (W) × 565 (D) × 500 (H) mm |
| Mass | 54 kg |
| Computer Requirements | Windows PC with available USB ports, Windows 10 64-bit or Windows 11 64-bit. Contact for additional details. |