Key Changes in USP <621>




        6. Adjustment of Chromatographic Conditions
                                  Key Changes in USP <621>
                                   -   Future Expectations
        Chromatography adjustments are outlined in Table 3.


           Parameter   Official from 01 Aug 2017 to 30 Nov 2022   Current USP : official as of 01 Dec 2022
           Elution Type   Isocratic   Gradient   Isocratic        Gradient
                                  Column parameters and flow rate
                                           Condition 1: No change of the substituent (e.g., no replacement of
                                           C18 by C8) and physicochemical characteristics of the stationary
                                           phase.
        Stationary Phase   No changes allowed
                                           Condition 2: A change from totally porous particles (TPP) columns to
                                           Superficially porous particles (SPP) columns allowed if condition 1
                                           requirements meet.
                    L and dP may be modified if the   L and dP may be modified if the ratio
        Column dimensions   ratio (L/dP) remains constant or  Changes not   (L/dP) remains constant or in the   Changes allowed. Same as
        (particle size, length)   in the range between -25% and  allowed   range between -25% and 50% of   Isocratic
                    50% of prescribed ratio.           prescribed ratio.  
                                  Key Changes in USP <621>
        Adjustments from   Other combination of L and dP
                                   -   Future Expectations
        totally porous to   allowed, provided plate number Changes not   Allowed, provided plate number (N)  Changes allowed. Same as

        superficially porous   (N) is within -25% to +50%   allowed       is within  -25% to +50%   Isocratic
                     Official from 01 Aug 2017 to 30 Nov 2022
        particles  Parameter   relative to prescribed column.    Current USP : official as of 01 Dec 2022   ♦  Where L= column  length;
                                   Changes not   Adjustment allowed if no change in  Changes allowed. Same as   dP=  particle  size’  F=flow;  dc=
        Internal diameter   Not mentioned
                                   allowed   dP and/or L    Isocratic         internal diameter of the column.
                                                            **Based on particle size (formula
                    *Based on particle size and   Changes in F,   *The flow rate of ±50% is permitted   Page 17 of 17   *When a change is made in
        Flow Rate   additionally ±50%  (formula for  dc and dP not  is adjusted for both the change in   for new gradient time calculation
                                                            from original gradient time, flow   particle  size  from  ≥3-µm  to
                    flow rate change)   allowed    column diameter and particle size
                                                            rate and the column dimensions)   <3-µm  an  additional  increase
        Column Temperature   ± 10°C         ± 10°C          ± 5°C              in  flow  rate  may  be  justified
                                       Mobile phase                            provided   column   efficiency
                                           Minor component comprises ≤
                    Minor component (specified as ≤50%)     Not mentioned      should  not  drop  by  >20%.
                                           100/n %)                          Similarly, when a change  is
                                           Adjustment allowed for minor
                    Adjustment allowed for minor components                    made  in particle  size  from
                                           components composition: ± 30 %
        Composition               Key Changes in USP <621>   Adjustment acceptable if principal
                    composition: ± 30 % (any component change   (any component change should not  peak elute within ± 15% of RT   <3-µm to ≥3-µm   an additional
                                   -   Future Expectations
                    should not exceed ±10 % absolute i.e., in                  reduction  in  flow  rate  may  be
                    relation to the total mobile phase)         exceed ±10 % absolute i.e., in   obtained with original condition
                                                  Current USP : officia
           Parameter   Official from 01 Aug 2017 to 30 Nov 2022   relation to the total mobile phase)   l as of 01 Dec 2022   justified  to  avoid  reduction  in
                                           Specified ratios 50:50, the mobile   column  efficiency  should  not
                    Specified ratios 50:50, the mobile phase ratio   phase ratio may be adjusted only
                    may be adjusted only within the range of 40:60–            drop by >20%.
                                           within the range of 40:60–60:40.
        Binary Mixture ratio   60:40.                       NA
                                           And for 2:98, the maximum allowed             2        2
                    And for 2:98, the maximum allowed adjustment  adjustment is within the range of   Page 17 of 17   *F2 = F1 × [(dc2  × dp1)/ (dc1 × dp2)]
                    is within the range of 1.4:98.6–2.6: 97.4.
                                           1.4:98.6–2.6: 97.4.
                                           Specified ratio 70:25:5 , the mixture   **When particle size changes
                    Specified ratio 60:35:5, the mixture ranges of             flow  rate  requires  changes
        Ternary Mixture ratio   50:45:5–70:25:5 or 58.5:35:6.5–61.5:35:3.5   ranges of 62.5: 32.5: 5 to 77.5:17.5:  NA    because   smaller   -particle
                                           5 or 68.5: 25: 6.5 to 71.5: 25: 3.5
                    would meet the requirement.
                                           would meet the requirement          columns will require  higher
        pH of aqueous   ±0.2 units, unless otherwise prescribed for both  ±0.2 units, unless otherwise prescribed for both isocratic and gradient   velocities  for  the  same
        component   isocratic and gradient     Key Changes in USP <621>        performance  (as measured by
                                   -   Future Expectations
        Concentration of salts in  Within ±10 % if permitted pH variation met,     Within ±10 % if permitted pH variation met, applies to both isocratic   reduced plate height).
        the buffer component   applies to both isocratic and gradient       and gradient.
        of MP  Parameter   Official from 01 Aug 2017 to 30 Nov 2022   Current USP : official as of 01 Dec 2022
                                                                                                2
                                                                               **tG2 = tG1 × (F1/F2) [L2 × dc ) / (L1 × dc 2 )]
                    Deviations not permitted. (For ±3nm at most                                2      1
        Detector wavelength   error verification should be done using detector  No adjustment permitted   No adjustment permitted   ♦♦ Where n= number of
                    manufacturer procedure)
                    Adjustment allowed for both isocratic and   ***Formula added for adjustment   components
                    gradient, as far as it is consistent with accepted  allowed upon column dimension   Page 17 of 17
        Injection volume                                    same as Isocratic   *** Vinj2 = Vinj1 (L2 dc ) / (L dc )
                                                                                              2
                                                                                                    2
                    precision, linearity, and detection limits. no   changed (formula NA for changes   2  1  1
                    formula mentioned.     from TPP columns to SPP column )
         NA: Not applicable
                             Table 3 : Chromatography adjustments
                                Table 3  Chromatography adjustments
         12  Spinco Biotech                                                                     April 2023
            Where L= column length; dP= particle size’ F=flow; dc= internal diameter of the column.
           *When a change is made in particle size from ≥3-µm to <3-µm an additional increase in flow rate may be justified provided column
           efficiency should not drop by >20%. Similarly, when a change is made in particle size from <3-µm to ≥3-µm   an additional reduction in
           flow rate may be justified to avoid reduction in column efficiency should not drop by >20%.
                  2
                        2
           *F2 = F1 × [(dc2  × dp1)/ (dc1  × dp2)]
                                                                      Page 17 of 17